When calibrating or testing a gauge or transducer with a pneumatic pressure controller you should protect the controller from any liquid contaminants that may be present in the device under test (DUT). A block and bleed valve (BBV) in-line with a coalescing filter is a good way to do this.
pressure gauge octa of containments is from the DUT to the controller. A BBV and a coalescing filter, oriented as observed in the diagram above, will to work to prevent contamination. The BBV is used to purge the majority of liquid from the DUT. The coalescing filter is a backup to get rid of any residual liquid which may be present even after the original purge and will separate entrained liquid droplets from gas flowing in direction of the controller. pressure gauge 10 bar is made to prevent harm to a pneumatic pressure controller from a contaminated DUT.
Purge the DUT of liquid containments
Before connecting the DUT to the system, remove as much liquid as possible from the DUT.
With the controller power off and both the block and bleed valve closed, connect the machine as shown above.
Use the controller to use a pressure near to the full scale pressure of the DUT.
Open the block valve slowly to pressurize the DUT (note: the gas will flow in the direction of the controller to the DUT ? ruthless to low pressure).
After the system is pressurized close the block valve.
Open the bleed valve and detox the system to atmospheric pressure. Take notice of the sump to see if any liquid is purged.
Close the bleed valve.
Repeat steps 3 to 7 until forget about liquid is observed appearing out of the system.
With diaphragm seal and the bleed valves securely closed, vent the controller.
After completion of the aforementioned procedure it will now be safe to calibrate the DUT. To calibrate, open the block valve and close the bleed valve. The coalescing filter will coalesce (join together) any liquid droplets which could still be present in the gas, and gravity will draw the liquid to the bottom of the filter housing. The filter will also remove particles larger than the specified size. An automatic or manual valve could be attached to underneath of the filter housing to periodically drain any liquid that has accumulated.
Note:
1. The coalescing filter won’t protect against a large slug of liquid that fills the housing and is forced through the filter membrane or liquid vapor.
2. The gas used to calibrate the DUT can flow in either direction within this system.
3. You will see a pressure drop over the filter in the flowing condition; at steady state pressure there is absolutely no pressure drop.